Resumen
The principle of the method specified is adding a caesium chloride solution to the sample as an ionization suppressant, aspirating the sample directly into the air/acetylene flame of an atomic absorption spectrometer and measuring the absorbance at a wavelength of 766,5 nm. Applies to water samples with a mass concentration of potassium in the range from 5 mg/l to 50 mg/l. This range can be extended if other dilution factors are chosen.
Informaciones generales
-
Estado: PublicadoFecha de publicación: 1993-05Etapa: Norma Internacional en proceso de revisión sistemática [90.20]
-
Edición: 1Número de páginas: 4
-
Comité Técnico :ISO/TC 147/SC 2ICS :13.060.50