Resumen
This document specifies procedures for the following:
a) primary particle-sizing calibration for particle sizes 1 µm(c) and larger, sensor resolution and counting performance of liquid automatic particle counters that are capable of analysing bottle samples;
b) secondary particle-sizing calibration using suspensions verified with a primary calibrated APC;
c) establishing acceptable operation and performance limits;
d) verifying particle sensor performance using a test dust;
e) determining coincidence and flow rate limits.
This document is applicable for use with hydraulic fluids, aviation and diesel fuels, engine oil and other petroleum-based fluids. This document is not applicable to particle-sizing calibration using NIST SRM 2806b primary calibration suspensions.
Informaciones generales
-
Estado: PublicadoFecha de publicación: 2022-03Etapa: Norma Internacional publicada [60.60]
-
Edición: 5Número de páginas: 51
-
Comité Técnico :ISO/TC 131/SC 6ICS :23.100.60
- RSS actualizaciones
Ciclo de vida
-
Anteriormente
RetiradaISO 11171:2020
-
Ahora