ISO 20411:2018
p
ISO 20411:2018
67923
недоступно на русском языке

ISO 20411:2018 specifies a method for determining the maximum count rate for an acceptable limit of divergence from linearity of the intensity scale in pulse counting magnetic sector-type secondary ion mass spectrometers or quadrupole secondary ion mass spectrometers. It uses a test based on depth profile analysis of two isotopes in a reference material which has a gradual concentration change between low and high concentration regimes. It also includes a correction method for saturated intensity caused by the dead time of the detector. The correction can increase the intensity range for 95 % linearity so that a higher maximum count rate can be employed for those spectrometers for which the relevant correction equations have been shown to be valid.

ISO 20411:2018 does not apply to time of flight mass spectrometers.

ISO 20411:2018 is only applicable to elements with minor isotopes. It is not applicable if the element is monoisotopic or contains isotopes with equal abundances.


Общая информация 

  •  :  Published
     : 2018-03
  •  : 1
  •  : ISO/TC 201/SC 6 Secondary ion mass spectrometry
  •  :
    71.040.40 Chemical analysis

Цели в области устойчивого развития

Данный стандарт разработан для достижения следующих Цель устойчивого развития:

Приобрести данный стандарт

ru
Формат Язык
std 1 92 PDF + ePub
std 2 92 Бумажный
  • CHF92

Жизненный цикл


Появились вопросы?

Ознакомьтесь с FAQ

Работа с клиентами
+41 22 749 08 88

Часы работы:
Понедельник – пятница: 09:00-12:00, 14:00-17:00 (UTC+1)

Будьте в курсе актуальных новостей ИСО

Подписывайтесь на наши новости, обзоры, а также на информацию о продуктах.