Reference number
ISO 17862:2022
International Standard
ISO 17862:2022
Surface chemical analysis — Secondary ion mass spectrometry — Linearity of intensity scale in single ion counting time-of-flight mass analysers
Edition 2
2022-09
Preview
ISO 17862:2022
80457
недоступно на русском языке
Опубликовано (Версия 2, 2022)

ISO 17862:2022

ISO 17862:2022
80457
Язык
Формат
CHF 96
Пересчитать швейцарские франки (CHF) в ваша валюта

Тезис

This document specifies a method for determining the maximum count rate for an acceptable limit of divergence from linearity of the intensity scale in single ion counting time-of-flight (TOF) secondary ion mass spectrometers using a test based on isotopic ratios in spectra from poly(tetrafluoroethylene) (PTFE). It also includes a method to correct for intensity nonlinearity arising from intensity lost from a microchannel plate (MCP) or scintillator and photomultiplier followed by a time-to-digital converter (TDC) detection system caused by secondary ions arriving during its dead-time. The correction can increase the intensity range for 95 % linearity by a factor of up to more than 50 so that a higher maximum count rate can be employed for those spectrometers for which the relevant correction formulae have been shown to be valid.

Общая информация

  •  : Опубликовано
     : 2022-09
    : Опубликование международного стандарта [60.60]
  •  : 2
  • ISO/TC 201/SC 6
    71.040.40 
  • RSS обновления

Жизненный цикл

Появились вопросы?

Ознакомьтесь с FAQ