Resumen
This document specifies a calibration procedure of the energy step and the energy scale for electron energy loss spectroscopy in (scanning) transmission electron microscopes to an uncertainty of ±3 % for the energy range 0 eV to 3 000 eV.
This document is intended for electron energy loss spectroscopy with transmitted electrons through sufficiently electron transparent samples, such as a thin foil sample, and is not designed for backscattered electrons from a bulk sample.
Informaciones generales
-
Estado: PublicadoFecha de publicación: 2022-07Etapa: Norma Internacional publicada [60.60]
-
Edición: 1Número de páginas: 15
-
Comité Técnico :ISO/TC 202/SC 3ICS :71.040.50